3-D Super-Resolution Ultrasound Imaging with a 2-D Sparse Array

Research output: Contribution to journalArticlepeer-review

81 Citations (Scopus)

Abstract

High frame rate 3-D ultrasound imaging technology combined with super-resolution processing method can visualize 3-D microvascular structures by overcoming the diffraction limited resolution in every spatial direction. However, 3-D super-resolution ultrasound imaging using a full 2-D array requires a system with large number of independent channels, the design of which might be impractical due to the high cost, complexity, and volume of data produced. In this study, a 2-D sparse array was designed and fabricated with 512 elements chosen from a density-tapered 2-D spiral layout. High frame rate volumetric imaging was performed using two synchronized ULA-OP 256 research scanners. Volumetric images were constructed by coherently compounding 9-angle plane waves acquired at a pulse repetition frequency of 4500 Hz. Localization-based 3-D super-resolution images of two touching sub-wavelength tubes were generated from 6000 volumes acquired in 12 seconds. In conclusion, this work demonstrates the feasibility of 3-D super-resolution imaging and super-resolved velocity mapping using a customized 2-D sparse array transducer.
Original languageEnglish
Pages (from-to)269-277
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
DOIs
Publication statusPublished - 1 Feb 2020

Fingerprint

Dive into the research topics of '3-D Super-Resolution Ultrasound Imaging with a 2-D Sparse Array'. Together they form a unique fingerprint.

Cite this