A reliable and inclusive method for assessing failure-causing mechanical wear in prosthetic feet

  • Michael A. Berthaume
  • , Sikandar Qayyum
  • , Ryley Ofoedu
  • , Taosif Mohammad
  • , Reza Hematti
  • , Nana Akosua Asamoa-Bonsu
  • , Someya Ali
  • , Gemma Ranson

Research output: Contribution to journal › Article › peer-review

5 Downloads (Pure)

Abstract

Wear assessment of prosthetic feet is critical for monitoring device performance and replacement, especially in rural or resource constrained areas where clinical follow-ups can be infrequent. Standardised, reliable methods for wear evaluation are limited. This study investigates the inter-observer reliability of a method for quantifying patterns and relative magnitudes of plantar surface wear of prosthetic feet and presents preliminary results for intra-observer error. Sixty-two plantar images of used SACH feet from Sri Lanka were scored by four independent groups a published method for quantifying prosthetic foot wear. A 20x10 grid was applied to each image, and wear was scored from 0-9 per cell based on visual indicators. Interobserver agreement was assessed on full (48 feet, 4 raters) and reduced (62 feet, 3 raters) datasets. Intraobserver reliability was tested by 2 raters re-scoring five images after 5 + days. Kendall's W evaluated agreement, and a modified scoring method was developed and evaluated. Interobserver agreement ranged from fair to excellent and was higher in the reduced dataset. Intraobserver agreement appears high, especially in central, less-worn areas of the foot. The modified method showed reduced intraobserver variability but lower interobserver agreement, largely due to increased ties affecting Kendall's W, not increased disagreement. The original and modified methods for wear qualification demonstrate high reliability, supporting their use as a low-cost, field-appropriate tool for monitoring prosthetic foot wear to enhance the evaluation and design of prosthetic feet, particularly in resource-limited settings. While the full method provides higher resolution data, the reduced method is more broadly applicable to different foot types. Clinically, this method offers a reliable, scalable approach to assessing prosthetic foot wear, potentially supporting community-based monitoring. Further testing and methodological alterations to encompass additional foot designs are required for clinical use.
Original languageEnglish
Article numbere0006545
JournalPLOS global public health
Volume6
Issue number5 May
DOIs
Publication statusPublished - 29 May 2026

Funding

Funding: We would like to thank In2STEM for providing students with the work placement and connecting the authors. We would further like to thank the JJCDR for help in collecting the feet. These organisations provided in-kind support by covering the salaries, overheads, etc., of employees who facilitated this research. Trips to Sri Lanka were funded in part by an NIHR PrOTeCTgrant (Award ID 16/137/45) award to Prof Anthony Bull (ImperialCollege London) and Research England funding awarded to MAB through London South Bank University (no award number). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

FundersFunder number
London South Bank University
National Institute for Health and Care ResearchID 16/137/45

    Cite this