Abstract
Contrast-enhanced ultrasound imaging creates one of the worst case scenarios for pulse compression due to depth and frequency dependent attenuation, high level of harmonic generation, phase variations due to resonance behavior of microbubbles, and increased broadband noise by microbubble destruction. This study investigates the feasibility of pulse compression with a matched filter in the existence of microbubbles with resonant behavior. Simulations and experimental measurements showed that the scattered pressure from a microbubble population excited by a chirp waveform preserves its chirp rate even for harmonic frequencies. Although, pulse compression by a matched filter was possible due to the conservation of the chirp rate, an increase on sidelobe levels were observed at fundamental and second harmonic frequencies. Therefore, using chirp excitation and a matched filter pair will increase the contrast-to-tissue ratio with a trade-off of decreased image quality.
© 2015 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
Original language | English |
---|---|
DOIs | |
Publication status | Published - 21 Oct 2015 |
Event | 2015 IEEE International Ultrasonics Symposium (IUS) - Duration: 21 Oct 2015 → … |
Conference
Conference | 2015 IEEE International Ultrasonics Symposium (IUS) |
---|---|
Period | 21/10/15 → … |
Keywords
- chirp coded excitation
- Microbubbles
- linear frequency modulation
- pulse compression
- matched filter