Skip to main navigation Skip to search Skip to main content

Non-Destructive Ultrasonic Estimation of Crystal Orientation Fabric Descriptors in Laboratory Ice-Core-Scale Specimens

Research output: Types of Thesis › PhD › peer-review

2 Downloads (Pure)

Abstract

Crystal orientation fabric (COF) governs the mechanical behaviour and deformation of polycrystalline ice and therefore influences ice-sheet flow dynamics. Existing characterisation methods, optical thin-section analysis and electron backscatter diffraction, provide direct crystallographic data but require destructive sample preparation and are limited to planar sections. No validated non-destructive framework exists for estimating directional fabric orientation within intact ice-core-scale specimens. This thesis develops a laboratory ultrasonic framework and validates its primary pulse–echo reflection pathway for estimating reduced in-plane fabric-sensitive orientation descriptors in intact ice-core-scale specimens. Three acquisition geometries are implemented: pulse–echo reflection as the primary validation pathway, opposed transmission as an independent geometry check, and multi-offset chord transmission as an exploratory effective-property mapping extension. Finite-difference timedomain modelling guides transducer geometry selection, defines bandpass-filtered envelope time-of-flight observables, and establishes a Cram´er–Rao lower bound of 1.7–2.2 for harmonic phase estimation under controlled simulation conditions. Pulse–echo reflection is validated across five disk specimens at three nominal frequencies (1.0–2.5 MHz) against registered 2-D optical references. Angular errors reach as low as 4 in the best case and four of five specimens fall within 20 at the lowest frequency tested (1.0 MHz). Three failure modes are identified and diagnostically characterised: grain-boundary scattering at high frequency, insufficient fabric concentration, and coordinate-frame registration error. Each produces a distinct diagnostic signature in harmonic amplitude, fit residual, and cross-frequency consistency. Opposed transmission on a single cylindrical specimen recovers fabric orientation within 10–15 of the optical reference, providing limited independent evidence that fabric-sensitive angular timing is not unique to the reflection geometry. Multi-offset chord inversion produces path-integrated effective-property maps with a physically consistent velocity ratio (v∥/v⊥ ≈ 1.09) across seven well-conditioned depth layers. The combined angular positioning and registration measurement floor is ±3.8 (95% coverage).
This measurement methodology combines systematic laboratory azimuthal ultrasonic scanning, second-order harmonic descriptor extraction, and criteria-filtered optical validation at ice-core scale. It defines the operating conditions, failure modes, and uncertainty limits of the primary pulse–echo descriptor-estimation pathway, while establishing secondary and exploratory routes for future transmission-based effective-property mapping. The method is applicable to ice-core laboratories and palaeoclimate research groups requiring repeated, non-destructive fabric characterisation of archived specimens without sectioning. Non-destructive access to fabric orientation enables the same core segment to be measured ultrasonically before and after mechanical testing or at multiple time points, which is not possible with existing destructive thin-section methods.
Original languageEnglish
QualificationDoctor of Philosophy
Awarding Institution
  • London South Bank University
Supervisors/Advisors
  • Lishman, Ben, Supervisor
  • Harput, Sevan, Supervisor
Award date26 Aug 2026
Publisher
DOIs
Publication statusPublished - 26 Aug 2026

Cite this